Blood Bank Refrigerator NBBR-200 offers a capacity of 658 liters for efficient blood storage. It uses a real forced air cooling system for rapid and uniform temperature control. This unit incorporates five types of audible and visual alarms for enhanced safety. It has two outer doors with heating and foam glass to prevent condensation. Our refrigerator includes a clear digital display for real-time monitoring of temperature and system status.
| Capacity (L/cu.ft) | 658/23.23 |
| Blood bags (450ml) | 360 units |
| Temperature range | 4±1℃ |
| Ambient temperature range | 10 to 32°C |
| Cooling type | Forced air cooling system |
| Defrost | Auto |
| Refrigerant | R134a |
| No. of shelves | 12 |
| No. of compressors | 1 |
| Compressor brand | SECOP |
| Sensor | NTC |
| Temperature controller | Microprocessor |
| Display | Display |
| High/low temperature alarm | Yes |
| Power off alarm | Yes |
| Backup battery malfunction alarm | Yes |
| Temperature controller malfunction alarm | Yes |
| Door ajar alarm | Yes |
| Power failure backup system alarm | 72 hours |
| Caster | Yes |
| Inner door | 6 |
| Outer door/type | 2/ Heating and foam glass doors |
| Test hole | 1/25mm |
| Thermal printer | Yes |
| Light | LED |
| Internal material | Stainless steel |
| External material | Color-sprayed steel |
| Insulation | PURF |
| Noise Level | 51.2 dB |
| Power consumption | 6.31 kWh/24h |
| Power supply | 220/50Hz |
| Current | 2.42A |
| Internal dimension | 1100×454×1325 mm |
| External dimension | 1220×648×1885 mm |
| Packaging dimension | 1300×735×2110 mm |
| Net weight | 202 kg |
| Gross weight | 257 kg |
Blood Bank Refrigerator NBBR-200 is used to store blood and blood components at a stable temperature to ensure their safety, potency, and compliance with medical standards. It is applied in blood banks, hospitals, research laboratories, and diagnostic centers.